- 25 -
11.7 Adjusting temperature limit values
If required, the circulator setting range of the operating temperature can be
limited.
This is not a safety element but a means to help avoid user faults when
operating the unit. The excess temperature protection must be set
separately.
Change to the function
H-L
with arrow buttons
7
in the menu
FU
. Then
press the enter button
8
to confirm the selection.
Press the menu button
3
:
The small display
5
shows
LL
(Low Limit temperature value).
Set the desired limit value with the arrow buttons
7
(the lowest possible
temperature is -30°C).
Press the enter button
8
.
The value is stored as low limit value
LL
.
Use the same procedure to set the high limit value
HL
(the highest possible
temperature is +100°C).
NOTE:
The new value is not saved until the enter key is pressed. If not
pressed, the circulator continues to use the old value.
Either press the menu button
3
again to quit the function mode and
return to display of the actual temperature, or the displays
5
and
6
will auto-
matically returns to actual temperature display mode after a short delay.