- 9 -
3. Set Up
3.1 Ambient conditions according to DIN EN 61010
• indoors, maximum. 2000 meters above sea level
• ambient temperature 5°C to 40° C
• relative humidity maximum. 80%/31°C (® 50%/40°C)
• excess voltage category II, contamination level 2
3.2 Downtime after transportation
(refrigerated circulators only)
Unfortunately, the unit is not always transported in an upright position so It is
possible lubrication oil leaked from the compressor into the cooling circuit.
The compressor may be damaged due to the lack of oil if the unit is started.
Do not operate the unit for 24 hours after setting up.
3.3 Ventilation
NOTE:
Keep all ventilation grids at the rear side free from obstruction to
ensure unhindered air circulation.
Blocked ventilation grids lead to increased unit heating which reduces
the unit's cooling capacity and hinders operation.
26
Ventilation Grids