- 21 -
11. Configuration
The unit is ready for operation after setting the desired set point temperatures
and adjusting the overtemperature protection. You can then adjust several
functions using the function mode
“FU”
.
Switch to the function mode by pressing the menu button
3
several times
(until display
5
shows
FU
).
Switch to the desired function
SOLL
,
RTA
,
CON
or
H-L
with the arrow
buttons
7
. Press the enter button
8
to confirm the selection made.
Move around through the submenus with the menu button
3
(see “Menu
Tree”).
Continue pressing the menu button
3
to quit the function mode and
returns to display of the actual temperature.
After a short delay, display
6
will automatically return to the actual temperature
display.
END is provided to skip the function mode without making any settings.
11.1 Set value S and fixed temperatures F1 to F3
The adjustment has already been made in 9.3 and 9.4. You can use the
function mode to adjust and change these set values
without
activating them
.
Press the arrow buttons
7
in the menu
FU
to change to the
SOLL
func-
tion. Press the enter button
8
to confirm this selection.
Press the menu button
3
to move through the submenus
S
,
F1
to
F3
(the
procedure for making adjustments is the same as in 9.3).
A correcting factor
c
should be associated with each value
(see 11.2).
7
6
3
8
5