- 17 -
9. Operation
9.1 Start
Turn the excess temperature protection to a setting above the desired
operating temperature
13
.
Start the circulator using the mains switch
1
.
This causes:
Display
6
to briefly show the device identification and then the version num
-
ber of the operating system software, e.g. “0.11”. Next the temperature at the
control sensor is displayed.
Display
5
shows which one of the four adjustable set values is currently acti-
vated (set temperature
“S”
or one of the fixed temperatures “
F1”,
“
F2”
,
”
F3"
(see 9.4).
The pump motor starts and circulates the bath liquid.
Cooling is switched on using switch
27
located on the front of the unit. The
compressor starts with a slight jerk. Cooling is activated only if cooling is
actually required.
9.2 Heating control lamp
Display
4
lights up when the heater is activated (set temperature is higher
than the actual temperature).
display
4
lights up constantly during the heating up phase,
display
4
flashes on and off during the control phase.
Display
4
does not light up if heating is not activated (set temperature is
lower than the actual temperature).
1
13
5
6
4
27