Installation

For packing purposes the Sample Concentrator is supplied unassembled in three parts; the base board, the column 
and carriage, and the gas chamber. The unit is re-assembled as follows:

1.  Unscrew the column fastening knob from the column support bracket already fixed to the base board.  
 

Locate the column on the two pins in the support bracket. Screw the fastening knob through the rear of the  

 

column and tighten firmly. 

2.  Align the threaded hole on the back of the gas chamber with the mounting on the carriage. Tighten the  
 

locking knob to secure the gas chamber in position.

3.  Connect the intake nozzle at the rear of the gas chamber to a suitable supply of gas via a pressure reducing  
 

valve. The gas pressure should be adjusted as required at the reducing valve but should not exceed 2 psi.  

 

Pressure in excess of 2 psi will cause gas to be vented and wasted.

4.  Place the Cole-Parmer Block Heater onto the base of the Sample Concentrator. It is positioned by means of  
 

two spigots on the base board which align with two centrally located screw heads on the underside of the  

 Dri-Block

®

 unit.

Carriage

Fastening knob

Location spigots

Needles

Base board

Gas chamber

Intake nozzle

Locking knob

Fastening knob

Column

4

Содержание EW-36620-40

Страница 1: ...Cole Parmer SC 200 Series Sample Concentrator Instruction Manual TECH0002 Version 1 1...

Страница 2: ...essing of Samples 8 Maintenance 8 Cleaning and Care 8 Carriage Friction 8 Replacing Sealing Pad 9 In Case of Contamination 9 Service and Repair 10 Warranty 10 Replacement Parts and Accessories 10 Diag...

Страница 3: ...into the test tubes The samples are heated from below by the heater and the flow of gas directed over the surface of the samples displaces the evaporated solvent from above the liquid s surface This s...

Страница 4: ...carriage Tighten the locking knob to secure the gas chamber in position 3 Connect the intake nozzle at the rear of the gas chamber to a suitable supply of gas via a pressure reducing valve The gas pr...

Страница 5: ...3506 you would place a needle in all the holes labelled C Only Sample Concentrators with the part number SC 200 96 or a replacement head part number 6106467 will have the 96 well array Tube size diame...

Страница 6: ...SC 200 matrix plate SC 200 96 matrix plate Inserting the Needles 6...

Страница 7: ...l misuse may reduce the effective life of the product and could cause a hazard Preventative maintenance should include keeping the product clean by protecting it from spillage contamination or corrosi...

Страница 8: ...he end of its service life the product must be accompanied by a Decontamination Certificate Servicing and Repair This product range does not require any routine servicing instructions for cleaning and...

Страница 9: ...ion F7209 Pack of 100 x 76mm long needles F7210 Pack of 100 x 127mm long needles FSC4NCS Pack of PTFE coated needles 100 x 76mm long FSC4NCL Pack of PTFE coated needles 100 x 127mm long 6101608 Spigot...

Страница 10: ...le diameter 6 35 mm Height 530 mm Width 240 mm Depth 295 mm Weight 3 5 Kg Customer and Technical Support For help and support contact Antylia Scientific Ltd Ltd Beacon Road Stone Staffordshire ST15 0S...

Страница 11: ...acid Ethanoic acid 118 44 Ethanol Alcohol 79 175 Ethyl acetate Ethyl ester 77 40 Heptane Dipropylmethane 98 120 Hexane n Hexane 69 335 Propan 2 ol Isopropanol 82 137 3 Methyl 1 butanol Isoamyl alcohol...

Страница 12: ...NOTES...

Страница 13: ...NOTES...

Страница 14: ...NOTES...

Страница 15: ...we cannot guarantee that interference will not occur in practise Where there is a possibility that injury damage or loss might occur if equipment malfunctions due to radio frequency interference or f...