Operation

Preparation

1.  Place the aluminium insert blocks to be used into the Dri-Block

®

 unit. The blocks should touch each other  

 

and be positioned centrally.

2.  Loosen the locking knob on the carriage. Remove the gas chamber, disconnecting the gas pipe if necessary,  
 

and lay the chamber upside down on a suitable work surface.

3.  Insert the hyperdermic needles in the pattern to suit the test tubes to be placed in the heater.

 

The needle guide holes are labelled with a letter corresponding with the type of insert block used in the heater. 

 

These are listed in the following table. As an example if you use the block for twelve 15mm test tubes (e.g.  

 

block number F3506) you would place a needle in all the holes labelled C.

 

Only Sample Concentrators with the part number SC-200-96 or a replacement head part number 6106467 

will 
 

have the 96 well array.

Tube size & diameter

Hole array

Hole label

5  x  6 deep

30

A

4  x  5 deep

20

B

3  x  4 deep

12

C

8 hole off-set array, e.g. F4490

D

 

Press each needle firmly, pointed end first, through the guide holes and through the sealing pad behind so  

 

that it is held firmly in position. Build up the desired pattern of needles to suit the arrangement of test tubes  

 

to be used. Do not put a needle in any position which will not have a corresponding sample to be concentrated

 

as this will only waste gas. The needles may be removed and replaced many times as the pad is self sealing.

4.   Replace the gas chamber onto the carriage and tighten the locking knob. Re-connect the gas pipe if necessary.

5

Содержание EW-36620-40

Страница 1: ...Cole Parmer SC 200 Series Sample Concentrator Instruction Manual TECH0002 Version 1 1...

Страница 2: ...essing of Samples 8 Maintenance 8 Cleaning and Care 8 Carriage Friction 8 Replacing Sealing Pad 9 In Case of Contamination 9 Service and Repair 10 Warranty 10 Replacement Parts and Accessories 10 Diag...

Страница 3: ...into the test tubes The samples are heated from below by the heater and the flow of gas directed over the surface of the samples displaces the evaporated solvent from above the liquid s surface This s...

Страница 4: ...carriage Tighten the locking knob to secure the gas chamber in position 3 Connect the intake nozzle at the rear of the gas chamber to a suitable supply of gas via a pressure reducing valve The gas pr...

Страница 5: ...3506 you would place a needle in all the holes labelled C Only Sample Concentrators with the part number SC 200 96 or a replacement head part number 6106467 will have the 96 well array Tube size diame...

Страница 6: ...SC 200 matrix plate SC 200 96 matrix plate Inserting the Needles 6...

Страница 7: ...l misuse may reduce the effective life of the product and could cause a hazard Preventative maintenance should include keeping the product clean by protecting it from spillage contamination or corrosi...

Страница 8: ...he end of its service life the product must be accompanied by a Decontamination Certificate Servicing and Repair This product range does not require any routine servicing instructions for cleaning and...

Страница 9: ...ion F7209 Pack of 100 x 76mm long needles F7210 Pack of 100 x 127mm long needles FSC4NCS Pack of PTFE coated needles 100 x 76mm long FSC4NCL Pack of PTFE coated needles 100 x 127mm long 6101608 Spigot...

Страница 10: ...le diameter 6 35 mm Height 530 mm Width 240 mm Depth 295 mm Weight 3 5 Kg Customer and Technical Support For help and support contact Antylia Scientific Ltd Ltd Beacon Road Stone Staffordshire ST15 0S...

Страница 11: ...acid Ethanoic acid 118 44 Ethanol Alcohol 79 175 Ethyl acetate Ethyl ester 77 40 Heptane Dipropylmethane 98 120 Hexane n Hexane 69 335 Propan 2 ol Isopropanol 82 137 3 Methyl 1 butanol Isoamyl alcohol...

Страница 12: ...NOTES...

Страница 13: ...NOTES...

Страница 14: ...NOTES...

Страница 15: ...we cannot guarantee that interference will not occur in practise Where there is a possibility that injury damage or loss might occur if equipment malfunctions due to radio frequency interference or f...

Страница 16: ...antylia com W coleparmer fr Italy T 39 0 284349215 E it sales antylia com W coleparmer it India T 9122 61394444 E info coleparmer in W coleparmer in China T 1 847 549 7600 E sales antylia com W colep...

Отзывы: