_____________________________________________________________________________
Cole-Parmer CLW-108i & CLW-128i Washers
12
cabinet hole toward the drain connection, then connect the drain
hose to your drain.
4)
CONNECT THE TAP AND DI WATER SUPPLIES
Connect and tighten all water connections securely.
5)
CONNECT THE ELECTRICAL SUPPLY
For a
Freestanding / Mobile
unit, roll the washer into its final
location and connect the supplied power cord to an appropriate
receptacle. Secure the unit by depressing the caster brake levers on
the unit’s casters. After circuit power is activated, the washer is
ready to operate.
For an
Undercounter
unit, connect the unit’s AC power plug to its
AC receptacle.
6)
TEST FOR LEAKS
Before final leveling and securement in position, test for leaks as
follows:
a.
Turn ON the electrical power and water supplies.
b.
Push the door closed until it is securely latched.
c.
Follow the “How to Start a Cycle” procedure on page 24
to select and operate a LIGHT cycle. This cycle executes
both tap water and DI water rinses.
d.
While the cycle is operating, carefully check all water
supply and drain connections to ensure that there are no
leaks. Correct any observed leaks.
7)
MOVE THE UNIT INTO FINAL POSITION
Using an assistant if necessary to take up excess cable and hose
slack, push the washer fully into its final position.