_____________________________________________________________________________
Cole-Parmer CLW-108i & CLW-128i Washers
7
INSTALLING THE WASHER
Your Cole-Parmer glassware washer must be enclosed on the top,
back, and both sides. The cabinet enclosure into which
Undercounter
units are installed supplies this protection.
Freestanding / Mobile
units are equipped with a #3 stainless steel
cabinet.
Only qualified, factory-trained personnel familiar with local plumbing
and electrical codes and ordinances should install the washer. For a
first-time installation, all site preparation work must be completed
before the washer can be installed with final utility connections. If
the new washer replaces an existing unit, verify that all plumbing
and electrical connections are properly sized and located.
To protect floor surfaces,
Undercounter
units are supplied with leg slides for
attachment to washer feet. These decrease
friction as the unit is moved across the floor.
Preparing for Installation
REQUIRED TOOLS
•
Phillips #2 screwdriver
•
Flat blade screwdriver
•
Adjustable wrenches
•
5/8" (16 mm) open-end wrench
•
Tape measure
•
Spirit level
•
Electric drill
•
Hole-saw bits up to 2" (50 mm) diameter
REQUIRED PLUMBING SUPPLIES
Your unit is supplied with two rear-mounted barbed fittings for
connection to your facility’s tap water and DI water supplies.
Two 6' lengths of 1/4" ID x 3/8" OD flexible tubing are provided.
Mating extension hoses are available in a Facility Installation Kit
(Part Number 445490). We recommend installing a shutoff valve
at the water supply connection. You will also need an appropriate
fitting for connecting the supplied drain hose to your drain system.
CAUTION
LEG
SLIDE
LEG
SLIDE