Cole-Parmer CLW-108i & CLW-128i Washers
8
REQUIRED UTILITIES
•
Electrical:
A 10-Amp rated outlet receptacle for a
dedicated, properly grounded branch circuit supplying
230V, 1
φ
, 50 Hz. Provide a local disconnect or switch for
safe maintenance access.
•
Tap Water:
18 – 60 PSI (124 – 414 kPa) @ minimum
3 gallons/min. (11 liters/min.).
Although not required, hot
water speeds water heat-up, decreasing cycle time.
To
facilitate service, install a shutoff valve (not supplied) in
the water supply line at a readily accessible location.
•
Deionized Water:
18 – 60 PSI (124 – 414 kPa) @
minimum 3 gallons/min (11 liters/min).
IMPORTANT
:
To prevent corrosion, D. I. Water should not exceed
1 M
Ω
/cm
resistivity.
•
Drain Supply:
3/4" (18 mm) Minimum ID piping.
DIMENSIONAL REQUIREMENTS
For all models, allow for 25"
(635 mm) door extension when
open. Allow at least 2" (5 cm)
space to the side of the door.
Prepare an installation space or
cabinet recess consistent with these
and the other dimensions listed for
your model in Table 1-2, “Washer
Dimensions,” on page 2.
Undercounter Model Only
— A typical cabinet recess for the
undercounter model is pictured here:
Rear connection locations are listed in Figure 4-1 below.
25"
(635 mm)
32-5/8" min.
(830 mm)
23-5/8"
(600 mm)
CAUTION
DOOR RADIUS:
25" (635 mm)