Models DC-1, -10, -20B DC Spark Testers
DC-1, -10, -20B Instruction Manual
–
Page 47
Taking Calibration Readings
4.
Turn on the DC-B and adjust the spark tester voltage to 3.00KV.
Record the voltage reading on the EVM.
5.
Repeat this step for all required test voltages in the current test
voltage range of the EVM.
6.
Turn off the spark tester and adjust the EVM range to for the
remaining readings. Repeat these steps for all required test voltages.
Note: Each time the Range is changed on the EVM the needle should be
adjusted to Zero.
Adjusting the Calibration
Entering the Calibration Mode
7.
Press and hold the Down Arrow Button and the CR Button for
approximately 4 seconds. The spark tester will enter the configuration
mode and disable the high voltage output.
8.
If the display is locked you will need to unlock the display with the PIN.
See “
” section for more information.
9.
Press the Down Arrow until “MCAL” is displayed on the voltage
display.
Adjusting the Spark Tester Output voltage
10.
Press the “B” button for several seconds until “
VOUT
” is displayed on
the voltage display.
11.
Press the “R” button. The Fault Count display should show “
S
.
00
.
0
”
representing the set point value.
12.
Using the UP/DOWN arrows adjust the output voltage up so that the Fault
Count display shows “
S
.
01
.
0
” for 1kV.
13.
Check the actual output voltage on the EVM, if this value is not within the rated
specification for the DC-B
Press the “CR” button. The Fault Count display
will
change to “
*
.
01
.
0
”.
14.
Using the UP/DOWN arrows adjust the reading on the EVM until the reading is
within the rated tolerance for the DC-B or as close to the value on the Fault
count display.
15.
Once the EVM and the value on the Fault Counter match press th
e “R” button to
redisplay the set point value “
S
.
01
.
0
”. Repeat these steps until all calibration
values are within tolerance.
Note: Calibration intervals begin at 1kV and are to be performed at 1kV increments.
16.
Press the “R” button until the Fault
Count display does not have the S or *.
Adjusting the Spark Tester Voltage Monitor
17.
Using the UP/DOWN arrows change the voltage display to read “VMON”
Содержание DC-1
Страница 2: ...Models DC 1 10 20B DC Spark Testers DC 1 10 20B Instruction Manual Page 1 Rev Null b0002 10 11 17 ...
Страница 7: ...Models DC 1 10 20B DC Spark Testers DC 1 10 20B Instruction Manual Page 6 Specifications ...
Страница 8: ...Models DC 1 10 20B DC Spark Testers DC 1 10 20B Instruction Manual Page 7 ...
Страница 58: ...Models DC 1 10 20B DC Spark Testers DC 1 10 20B Instruction Manual Page 57 ...