Models DC-1, -10, -20B DC Spark Testers
DC-1, -10, -20B Instruction Manual
–
Page 36
Any Barewire Alarm (ABW)
This output will operate in 2 different modes,
“NLC”
,
”LCH”
. The Any Barewire
alarm will respond only to a Metal Contact or Gross Barewire fault type. See the
“Fault Typing” section of the manual for more information.
Any Barewire Alarm Mode (ABMO)
–
Non Latch Mode
The output will change state
for the configured “
ABMS
” time
.
–
Latch Mode
The output will change state until it is reset at the front panel or
remotely.
Any Barewire Alarm Time (ABMS)
Set in seconds (.01
–
5.0). If
“ABMO”
is set to
“NLC”
, this is the time for
which the output will change state.
Any Pinhole Alarm (APH)
This output will operate in 2 different modes,
“NLC”
,
”LCH”
. The Any Pinhole
alarm will respond only to a Pinhole or Multi pinhole fault type.
See the “Fault
Typing” section of the manual for more information.
Any Pinhole Alarm Mode (APMO)
–
Non Latch Mode
The output will change state
for the configured “
APMS
” time
.
–
Latch Mode
The output will change state until it is reset at the front panel or
remotely.
Any Pinhole Alarm Time (APMS)
Set in seconds (.01
–
5.0). If
“APMO”
is set to
“NLC”
, this is the time for
which the output will change state.
Содержание DC-1
Страница 2: ...Models DC 1 10 20B DC Spark Testers DC 1 10 20B Instruction Manual Page 1 Rev Null b0002 10 11 17 ...
Страница 7: ...Models DC 1 10 20B DC Spark Testers DC 1 10 20B Instruction Manual Page 6 Specifications ...
Страница 8: ...Models DC 1 10 20B DC Spark Testers DC 1 10 20B Instruction Manual Page 7 ...
Страница 58: ...Models DC 1 10 20B DC Spark Testers DC 1 10 20B Instruction Manual Page 57 ...