Models DC-1, -10, -20B DC Spark Testers
DC-1, -10, -20B Instruction Manual
–
Page 33
Pinhole Alarm (PHA)
This output will operate in 2 different modes,
“NLC”
,
”LCH”
. The Pinhole alarm
will respond only to a
‘Pinhole’ fault type
.
See the “New Features” section of
the manual for more information on fault typing.
Pinhole Alarm Mode (PHMO)
–
Non Latch Mode
The output will change state
for the configured “
PHMS
” time
.
–
Latch Mode
The output will change state until it is reset at the front panel or
remotely.
Pinhole Alarm Time (PHMS)
Set in seconds (.01
–
5.0). If “PHMO” is set to “NLC”,
this is the time for
which the output will change state.
Metal Contact Alarm (MCA)
This output will operate in 2 different modes,
“NLC”
,
”LCH”
. The Metal Contact
alarm will respond only to a
‘direct
metal contact
’ fault type
.
See the “Fault
Typing” section of the
manual for more information.
Metal Contact Alarm Mode (MCMO)
–
Non Latch Mode
The output will change state
for the configured “
MCMS
” time
.
–
Latch Mode
The output will change state until it is reset at the front panel or
remotely.
Metal Contact Alarm Time (MCMS)
Set in seconds (.01
–
5.0). If “MCMO” is set to “NLC”,
this is the time for
which the output will change state.
Multi Pinhole Alarm (MPA)
This output will operate in 2 different modes,
“NLC”
,
”LCH”
. The Multi Pinhole
alarm
will respond only to a ‘Multi Pinhole’ fault type
.
See the “Fault Typing”
section of the manual for more information.
Содержание DC-1
Страница 2: ...Models DC 1 10 20B DC Spark Testers DC 1 10 20B Instruction Manual Page 1 Rev Null b0002 10 11 17 ...
Страница 7: ...Models DC 1 10 20B DC Spark Testers DC 1 10 20B Instruction Manual Page 6 Specifications ...
Страница 8: ...Models DC 1 10 20B DC Spark Testers DC 1 10 20B Instruction Manual Page 7 ...
Страница 58: ...Models DC 1 10 20B DC Spark Testers DC 1 10 20B Instruction Manual Page 57 ...