Models DC-1, -10, -20B DC Spark Testers
DC-1, -10, -20B Instruction Manual
–
Page 43
Fieldbus Spark Tester Parameter Addresses
The fieldbus parameters are 2 bytes each. The LSB will be at the lower byte offset, the MSB at the higher byte offset. As
an example the setpoint is 5000 volts (1388H in hexadecimal). The setpoint is at a word address 6. When reading the
two bytes of the setpoint the LSB (value of 88H) would be received first and the MSB (value of 13H) would be received
second. Some PLCs start with addresses of 0, some start with 1. The table below assumes the addresses start with 0.
To check the addressing, read from location 103 and you should read back a value of 15000.
Command
Address R/W
Notes
RESERVED
0
-
RESERVED
1
-
REMOVE VOLTAGE COMMAND
2
W
Will turn off/on the voltage without changing the setpoint
AF RESET COMMAND
3
W
Will reset All Fault Alarms
REMOTE ENABLE
4
RW
Enables Data writes from External bus
SETPOINT VOLTS
5
RW
Requested voltage in Volts
ACTIVE LINE SPEED
6
RW
Line Speed in ft/m
VMON VOLTS
7
R
Voltage at metering windings (Actual Voltage) in Volts
EVMON VOLTS
8
R
Voltage at Electrode in Volts
IMON PERCENT LOAD
9
R
Percent of maximum load on the system
AF COUNT
10
RW
All fault count
PH COUNT
11
RW
Pinhole fault count
MC COUNT
12
RW
Metallic Contact fault count
MP COUNT
13
RW
Multi-pinhole fault count
GB COUNT
14
RW
Gross Bare Wire fault count
STATUS FLAGS 0
15
R
See Definitions below
STATUS FLAGS 1
16
R
See Definitions below
Below are the individual bits for Status Word 0
GLOBAL VOLTAGE ENABLED
Bit 0
R
Set when all conditions allow voltage to be generated
AV HVL STATUS
Bit 1
R
Set when the Actual Voltage >= HVL On voltage
AV WDL STATUS
Bit 2
R
Set when voltage is in the range of the Watchdog Low and High levels
FP STATUS
Bit 3
R
Set when the fault is detected and cleared at end of fault pulse duration
AF ALARM STATUS
Bit 4
R
State of the AF Alarm cleared by Fault Reset
PH ALARM STATUS
Bit 5
R
State of the PH Alarm cleared by Fault Reset
MC ALARM STATUS
Bit 6
R
State of the MC Alarm cleared by Fault Reset
MP ALARM STATUS
Bit 7
R
State of the MP Alarm cleared by Fault Reset
GB ALARM STATUS
Bit 8
R
State of the GB Alarm cleared by Fault Reset
AF COUNT LIMIT STATUS
Bit 9
R
Set if AF Count >= AF Count Limit cleared by Fault Count Reset
PH COUNT LIMIT STATUS
Bit 10
R
Set if PH Count >= PH Count Limit cleared by Fault Count Reset
MC COUNT LIMIT STATUS
Bit 11
R
Set if MC Count >= MC Count Limit cleared by Fault Count Reset
MP COUNT LIMIT STATUS
Bit 12
R
Set if MP Count >= MP Count Limit cleared by Fault Count Reset
GB COUNT LIMIT STATUS
Bit 13
R
Set if GB Count >= GB Count Limit cleared by Fault Count Reset
FAULT CO LIMIT STATUS
Bit 14
R
Set if Fault Co Count >= Fault Co Count Limit cleared by Fault Count Reset
AVSP WD PERCENT STATUS
Bit 15
R
Set when Actual Voltage is within the AVSP Percent of the Setpoint
Below are the individual bits for Status Word 1
EVSP WD PERCENT STATUS
Bit 0
R
Set when Electrode Voltage is within the EVSP Percent of the Setpoint
ABW ALARM STATUS
Bit 1
R
Indicates a Bare Wire fault has been detected cleared by Fault Reset
APH ALARM STATUS
Bit 2
R
Indicates a Pinhole fault has been detected cleared by Fault Reset
UNIT ERROR STATUS
Bit 7
R
Indicates the system has detected an error
GPO 0 STATUS
Bit 8
R
State of General Purpose Output 0
GPO 1 STATUS
Bit 9
R
State of General Purpose Output 1
GPO 2 STATUS
Bit 10
R
State of General Purpose Output 2
GPO 3 STATUS
Bit 11
R
State of General Purpose Output 3
GPI 0 STATUS
Bit 12
R
State of General Purpose Input 0
GPI 1 STATUS
Bit 13
R
State of General Purpose Input 1
GPI 2 STATUS
Bit 14
R
State of General Purpose Input 2
GPI 3 STATUS
Bit 15
R
State of General Purpose Input 3
Содержание DC-1
Страница 2: ...Models DC 1 10 20B DC Spark Testers DC 1 10 20B Instruction Manual Page 1 Rev Null b0002 10 11 17 ...
Страница 7: ...Models DC 1 10 20B DC Spark Testers DC 1 10 20B Instruction Manual Page 6 Specifications ...
Страница 8: ...Models DC 1 10 20B DC Spark Testers DC 1 10 20B Instruction Manual Page 7 ...
Страница 58: ...Models DC 1 10 20B DC Spark Testers DC 1 10 20B Instruction Manual Page 57 ...