Models DC-1, -10, -20B DC Spark Testers
DC-1, -10, -20B Instruction Manual
–
Page 35
Pinhole Limit (PHL)
This output will change state
when the “Pinhole”
count is equal to or greater
than the configured
“PHLT”
.
Pinhole Count Limit (PHLT)
Amount of counts allowed before the output responds.
Metal Contact Limit (MCL)
This output will change state
when the “Metal Contact”
count is equal to or
greater than the configured
“MCLT”
.
Metal Contact Count Limit (MCLT)
Amount of counts allowed before the output responds.
Multi Pinhole Limit (MPL)
This output will change state
when the “Multi Pinhole”
count is equal to or
greater than the configured
“MPLT”
.
Multi Pinhole Count Limit (MPLT)
Amount of counts allowed before the output responds.
Gross Barewire Limit (GBL)
This output will change state
when the “Gross Barewire”
count is equal to or
greater than the configured
“GBLT”
.
Gross Barewire Count Limit (GBLT)
Amount of counts allowed before the output responds.
Fault Combination Limit (FCL)
–
Must be configured via USB computer interface
This output will change state when the conditions of the
‘COLT’
are met.
Fault Combination Limit Mode (COLT)
“0”
: The output will change state when either the Pinhole Count, or the
Multi Pinhole Count are equal to or greater than
the ‘’MPLT” or “PHLT”
.
“1”
: The output will change state when either the Metal Contact count,
or the Gross Bare Wire Count are equal to or greater than
‘GBLT’ or
‘MCLT’
Содержание DC-1
Страница 2: ...Models DC 1 10 20B DC Spark Testers DC 1 10 20B Instruction Manual Page 1 Rev Null b0002 10 11 17 ...
Страница 7: ...Models DC 1 10 20B DC Spark Testers DC 1 10 20B Instruction Manual Page 6 Specifications ...
Страница 8: ...Models DC 1 10 20B DC Spark Testers DC 1 10 20B Instruction Manual Page 7 ...
Страница 58: ...Models DC 1 10 20B DC Spark Testers DC 1 10 20B Instruction Manual Page 57 ...