Digital Color Doppler Palm Ultrasound System
56
To avoid electrical shock, always turn off the system and disconnect the system before cleaning the
system.
CAUTION
Take extra care when handling the lens face of the system. The lens face is especially
sensitive and can easily be damaged by rough handling. NEVER use excessive force when
cleaning the lens face.
Rinse the system with enough clean potable water to remove all visible soap residue.
Air dry or dry with a soft cloth.
CAUTION
To minimize the risk of infection from blood-borne pathogens, you must handle the system and
all disposables that have contacted blood, other potentially infectious materials, mucous
membranes, and non-intact skin in accordance with infection control procedures. You must
wear protective gloves when handling potentially infectious material. Use a face shield and
gown if there is a risk of splashing or splatter.
Disinfecting the systems:
After each use, please disinfect the systems.
Ultrasound systems can be disinfected using liquid
chemical germicides. The level of disinfection is directly related to the duration of contact with the germicide.
Increased contact time produces a higher level of disinfection.
In order for liquid chemical germicides to be effective, all visible residue must be removed during the
cleaning process. Thoroughly clean the system, as described earlier before attempting disinfection.
You MUST disconnect the system from the displayer prior to cleaning/disinfecting the system. Failure to do
so could damage the system.
DO NOT soak systems in liquid chemical germicide. Soaking may cause system damage and early failure of
the enclosure, resulting in possible electric shock hazard.
Prepare the germicide solution according to the manufacturer's instructions. Be sure to follow all
precautions for storage, use and disposal. The transducer is not designed to be totally submerged in
fluid. Permanent damage will result if the entire transducer is submerged.
Place the cleaned and dried system in contact with the germicide for the time specified by the germicide
manufacturer.
Ultrasound transducers can easily be damaged by improper handling and by contact with certain chemicals.
Failure to follow these precautions can result in serious injury and equipment damage.