![Chison SonoEye P2 Скачать руководство пользователя страница 58](http://html.mh-extra.com/html/chison/sonoeye-p2/sonoeye-p2_user-manual_2602726058.webp)
Digital Color Doppler Palm Ultrasound System
53
WARNING
The systems are designed to be used only with this ultrasound system. Use of these systems
on any other device or a non-qualified device may cause electrical shock or damage on the
system/transducer.
10.2.2 Cleaning and Disinfecting
This section provides information and instructions for properly cleaning and disinfecting the system.
Following these instructions will also help to avoid damaging the system during cleaning and disinfection.
After each exam, clean and disinfect the system.
CAUTION:
Prevent any fluid from entering electrical or metal portions of the cable’s connector
during the cleaning and disinfecting process. Damage due to fluid in these areas may result.
Scrub the system as needed using a soft sponge, gauze, or cloth to remove all visible residue from
the system surface.
Use a soft cloth to clean the cable and the user section of the system with the cleaning disinfectant
liquid. Make sure that the surface of the system and cable is wetted thoroughly with the cleaning-disinfectant.
Allow system to air dry completely.
Recommended materials for cleaning and disinfecting transducer
Solution
Origin
Qualified
Use
Active Ingredients
Purpose
Time
WIP'ANIOS
FRA
Wipe
Isopropyl
alcohol/Didecyl
dimethyl ammonium
chlorides
Disinfection
NA
80% Ethanol
(
Using clean
cloth soak the liquid
)
Any
Wipe
Ethanol
Clean or
Disinfection
NA
70% Isopropyl alcohol
(
Using clean cloth soak
the liquid
)
Any
Wipe
Isopropyl alcohol
Clean or
Disinfection
NA
Clinell Sporicidal Wipes
GBR
Wipe
Sodium
Percarbonate Citric
Acid
Disinfection
NA
mikrozid PAA wipes
DEU
Wipe
peracetic
acid,Hydrogen
peroxid,Acetic acid
Disinfection
NA
Universal wipes
GBR
Wipe
Quaternary
ammonium
Disinfection
NA
perform classic wipes EP
DEU
Wipe
EP
Disinfection
NA
CAUTION
These transducers are not designed to withstand heat sterilization methods. Exposure to
temperatures in excess of 60 º C will cause permanent damage. The transducers are not
designed to be totally submerged in fluid, as permanent damage will result if the entire
transducer is submerged.
Содержание SonoEye P2
Страница 30: ...Digital Color Doppler Palm Ultrasound System 25 ...
Страница 74: ...Digital Color Doppler Palm Ultrasound System 69 ...