ECO 6 Digital Color Doppler Ultrasound System
185
Appendix F: Transducer Maximum Surface Temperature
According to the requirements of the section
201.11
in the standard IEC/EN 60601-2-37:2007, the
transducer surface temperature has been tested in two kinds of conditions
: the transducer suspended in
still air or transducer contacting human-tissue mimicking material. The calculation of the expanded uncertainty
is based on the ISO Guide tout ye Expression of uncertainty in measurement. Three transducer samples have
been tested and the confidence coefficient is at 95%,the value of t.975 is 4.30.
The measurement data were obtained under the test conditions employed at CHISON.
Transducer
model
Maximum
surface
temperature(
℃
)
C3-A
<41.0
MC6-A
<41.0
L7M-A
<41.0
R7-A
<41.0
L7S-A
<41.0
MC3-A
<41.0
V6-A
<41.0
P3-A
<41.0