![Chison ECO 6 Скачать руководство пользователя страница 175](http://html.mh-extra.com/html/chison/eco-6/eco-6_operation-manual_2602723175.webp)
ECO 6 Digital Color Doppler Ultrasound System
162
Acoustic Output Table
System: ECO 6
Transducer Model: MC6-A
Operating Mode: B+M
Index Label
MI
TIS
TIB
TIC
Scan
Non-scan
Non-scan
Aaprt≤1c
m
2
Aaprt
>
1cm
2
Global Maximum
Index Value
0.3
0.4
0.1
#
Associated
p
ra
Mpa
0.61
Acoustic
P
mW
16
16
#
Parameter
min of [P
α
(Z
s
),
I
ta.α
(Z
s
) ]
mW
Z
s
cm
Z
bp
cm
Z
b
cm
1.95
Z at max.I
pi.α
cm
2.05
d
eq
(Z
b
)
cm
3.39
f
awf
MHz
5.56
5.46
5.46
#
Dim of
A
aprt
X
cm
2.43
2.43
#
Y
cm
0.7
0.7
#
Other
t
d
μs
0.25
Information prr
Hz
400
p
r
at max.I
pi
MPa
0.75
d
eq
at max.I
pi
cm
3.27
I
pa.α
at max.MI
W/cm
2
12.25
Operating
Mode
B+M
B+M
B+M
#
Control
Focus
cm
3
5
5
#
Conditions
A Power
%
100
100
100
#
NOTE1: Data should only be entered in one of the columns related to TIS.
NOTE2: Information need not be provided regarding TIC for any TRANSDUCER ASSEMBLY not intended
for transcranial or neonatal cephalic uses.
NOTE3: If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns related
to TIS, TIB or TIC.
NOTE4: If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the column related
to MI.
WWW.CFS.IT - CFS PRODOTTI MEDICALI SRL
Содержание ECO 6
Страница 2: ...WWW CFS IT CFS PRODOTTI MEDICALI SRL ...