9.3
Probe Operation Instructions
Instructions for connecting, activating, deactivating, disconnecting, transporting and storing the probes.
9.3.1
Scanning the Patient
In order to assure optimal transmission of energy between the patient and probe, a conductive gel must be
applied liberally to the patient where scanning will be performed.
After the examination is complete, follow the cleaning and disinfecting, or sterilizing procedures as
appropriate.
9.3.2
Operating Transvaginal probe
The transvaginal probe is an endo-cavity probe, for the operation safety, please refer to “Care and
Maintenance” for cleaning and disinfection.
The temperature at the tip of the probe displays on the screen for monitoring. No temperature above 43º C is
allowed. It also depends on the patient’s body temperature. When the temperature of probe tip exceeds 43º C,
the probe will stop working immediately to protect the patient.
Transvaginal probe should be used with FDA approved condom or probe cover. See the following instructions
to put the probe into the condom:
CAUTION
Some patients may be allergic to natural rubber or medical device with rubber contains. FDA suggests that
the user identify these patients and be prepared to treat allergic reactions promptly before scanning.
Only water-solvable solutions or gel can be used. Petroleum or mineral oil-based materials may harm the
cover.
When the transvaginal probe is activated outside patient’s body, its acoustic output level should be
decreased to avoid any harmful interference with other equipment.
Operation Procedure: