ECO 6 Digital Color Doppler Ultrasound System
172
Acoustic Output Table
System: ECO 6
Transducer Model: P3-A
Operating Mode: B
Index Label
MI
TIS
TIB
TI
C
Scan
non-scan
non-scan
A
aprt
≤1
A
aprt
>1
Global Maximum
Index Value
1.2
1.7
Ass
ocia
ted
Aco
ustic
Para
met
er
p
r.3
p
r,α
(MPa)
1.84
W
o
P
(mW)
269.95
min of
[W
.3
(z
1
),
I
TA.3
(z
1
)]
min of
[Pα(Zs),Ita,α(Z
s)]
(mW)
z
1
Z
s
(cm)
z
bp
Z
bp
(cm)
z
sp
Z
b
(cm)
Z
@PII
.3max
Z at max I
pi α
(cm)
4.5
d
eq
(z
sp
)
d
eq
(Z
b
)
(cm)
f
c
f
awf
(MHz)
2.28
2.38
Dim of A
aprt
Dim of A
aprt
X (cm)
1.92
Y (cm)
1.58
Oth
er
Info
rmat
ion
PD
t
d
(µsec)
0.97
PRF
prr
(Hz)
3337.5
p
r
@PII
max
p
r
at max I
pi
(MPa)
3.83
d
eq
@PII
max
d
eq
at max I
pi
(cm)
Focal Length
Focal Length
FL
x
(cm)
0.32
FL
y
(cm)
0.24
I
PA.3
@ MI
max
I
pi
at max MI
(W/cm
2
)
98.54
Ope
ratin
g
Con
trol
Con
ditio
ns
Mode
Mode
NA
B
B
Focus
Focus
(cm)
4
1
Power
Power
(%)
100
100
Notes:
(a) This index is not required for this operating mode. see section 4.1.3.1. of the Output Display Standard
(NEMA UD-3).
(b) This probe is not intended for transcranial or neonatal cephalic uses.
(c) This formulation for TIS is less than that for an alternate formulation in this mode.
# No data are reported for this operating condition since the global maximum index value is not reported for the
reason listed.