ECO 6 Digital Color Doppler Ultrasound System
114
Cleaning procedure:
1. Wipe the remaining coupling gel and blot with flowing clear water. Avoid the joint part between cable and
probe touching the water or others.
2. Use wet gauze or other soft cloth with a little liquid soap to clean the probe totally. Don’t use cleaning agent
and cleaner with abrasiveness.
3. Use flowing water to rinse fully. Use soft cloth which has been soaked by the concentration of 70%
isopropyl alcohol to scrub. Then check the probe to make sure there is no blot.
4. Use clean cloth to dry the probe.
Caution:
Don’t put the body surface probe into liquid below acoustical lens. Intracavity probe can’t
exceed insertion region. Prohibit putting connector of probe into any liquid.
Disinfection
Disinfection procedure is fit for intracavity probe.
If it is necessary to use in surgery, please abide with instruction of professional disinfection person.
Disinfection procedure:
1. Obey the cleaning procedure to clean the probe totally.
2. Prepare and retreat the disinfection solution according to the instruction of manufacturer.
3. Put the insertion region of probe into disinfection solution, the inserting depth can’t exceed insertion region.
4. Don’t let the probe connector touch any liquid.
5. Soak the probe for 3 h.
6. Pull out the probe, immediately rinse totally with sterile water and saline. To make sure of no any solution
remaining. Please obey the rule to do right rinsing procedure including enough rinsing water and times.
7. When probe is used in sterile area, make sure to use disposable sterile probe hood.
Caution:
1. Don’t soak probe connector into any liquid.
2. Don’t let the soaking depth of intracavity probe exceed insertion region.
3. Prohibit soaking the probe in the liquid for more than 12 h.
4. Only use qualified disinfection solution.
Deposit:
Please replace the probe in clean and dry environment, avoid direct sunshine.
Keep the environment to deposit the probe during -10
-
50
℃
, Do not put it in high pressure and vacuum
environment.
When accessing probe be careful and avoid ruin.
During transportation or leisure, the probe should be deposited in probe box.
8.3
Safety Check
To make sure this device works normally, a piece of maintenance plan is suggested to make to check the
device regularly If any abnormal phenomena, immediately stop operation and contact with Sale Office,