ECO 6 Digital Color Doppler Ultrasound System
3
2.2
Electrical Safety
Type of protection against electric shock
Class I Equipment
CLASS I EQUIPMENT in which protection against electric shock does not rely on BASIC INSULATION
only, but includes a protective earth ground. This additional safety precaution prevents exposed metal parts
from becoming LIVE in the event of an insulation failure.
NOTE
: The mains supply shall be cut off after disconnecting the power line and the net power.
Degree of protection against electric shock
Type BF Applied part
(for Probes marked with BF symbol)
TYPE BF APPLIED PART providing a specified degree of protection against electric shock, with particular
regard to allowable LEAKAGE CURRENT
Level of protection against harmful ingress of water
Parts of probe likely to come into contact with operator or patient meet the requirements of drip-proof
equipment (IPX1)
Parts of probe intended to be immersed in normal use meet the requirements of watertight equipment (IPX7)
The IP Classification of System is Ordinary Equipment (IPX0)
Safety level when used in the presence of FLAMMABLE ANAESTHETIC
MIXED WITH AIR (or WITH OXYGEN or WITH NITROUS OXIDE):
The Equipment is not suitable for use in the environment with FLAMMABLE ANAESTHETIC MIXED
WITH AIR (or WITH OXYGEN or WITH NITROUS OXIDE)
Mode of operation
Continuous Operation
For maximum safety, always follow these guidelines:
Proper grounding of the system is critical to avoid electrical shock. For protection, ground the chassis with
a three-wire cable and plug, and plug the system into a hospital-grade, three-hole outlet.
Do not remove or circumvent the grounding wire.
Do not remove the protective covers on the system. These covers protect users from hazardous voltages.
Cabinet panels must remain in place while the system is in use. A qualified electronic technician must make all
internal replacements.
Do not operate this system in the presence of flammable gases or anesthetics.
All peripheral devices (unless certified as medical grade) that are connected to the system must be powered
through the electrical outlet through an optional isolation transformer.