ECO Series VET Digital Color Doppler Ultrasound System
23
Load and unload the system to a vehicle parked on a level surface.
Load the unit abroad the vehicle carefully and over its center of gravity. Keep the unit still and upright.
Ensure that the transporting vehicle can bear the weight of system plus the passengers.
Secure the system firmly with straps or as directed within the vehicle to prevent movement during transport.
Any movement, coupled with the weight of the system, could cause it to break loose.
Drive carefully to prevent damage from vibration. Avoid unpaved roads, excessive speeds, and erratic stops
or starts.
3.5.2
Powering the System
3.5.2.1
Acclimation Time
After being transported, the unit requires one hour for each 2.5 º
increment if its temperature is below 10 º
C or
above 40 º
C.
NOTE
Please keep at least 20 to 30 cm spare space away from the back of the system to ensure well ventilation.
Otherwise, with the increasing of the temperature inside the unit, malfunction may occur.
3.5.2.2
Connecting the electric power
After making sure that the AC power supply in hospital is in normal status, and this AC voltage type matches
to the power requirements indicated on the label of system, then please connect the plug of power cord to the
DC IN socket at the rear panel of the system, and connect the other end of power cord to the AC power supply
socket in hospital.
Please use the power cable provided by the manufacturer, other type of power cable is not allowed.
Press
1 s, system boot,
Press
and pop up dialog for shut down. Click the enter key then power off.
Or press
3 s, system will forced shut down
CAUTION
Connecting the system to the wrong AC power supply may cause damage to the system and danger to the
operators and animals.
3.5.3
Probe Installation
Caution
:
Please only use the probes provided by manufacturer for this model, other types of probes
are not allowed to use with this system! Otherwise it may cause the damage to the system and the probe.
CAUTION
Before connecting the probe, please carefully check the probe lens, probe cable and probe connector to see
whether there is anything abnormal, such as cracks, falls off. Abnormal probe is not allowed to connect to the
system; otherwise there is possibility of electricity shock.
Hold the probe connector lock switch, and insert the connector socket vertically.