![Chison ECO 1VET Скачать руководство пользователя страница 150](http://html.mh-extra.com/html/chison/eco-1vet/eco-1vet_operation-manual_2602722150.webp)
ECO Series VET Digital Color Doppler Ultrasound System
135
Transducer Model: L7S-A
Operating Mode: PW
Index label
MI
TIS
TIB
TIC
At
surface
Below
surface
At
surface
Below
surface
Maximum index value
0.80
1.20
1.70
–
Index component value
1.10
1.20
1.60
1.70
Acoustic
Parameters
p
r,α
at Z
MI
Mpa
2.26
P
mW
30.00
30.00
–
P
1x1
mW
11.72
11.72
Z
s
cm
1.44
Z
b
cm
1.50
Z
MI
cm
1.46
z
pii,α
cm
1.50
f
awf
MHz
7.95
8.15
7.97
–
Other
Information
prr
Hz
3424.70
srr
Hz
–
n
pps
1
I
pa,α
at z
pii,α
W/cm
2
141.73
I
spta,α
at z
pii,α
or
z
sii,α
mW/cm
2
142.23
I
spta
at z
pii
or z
sii
mW/cm
2
160.35
p
r
at z
pii
Mpa
3.05
Operating
Control
Conditons
Focus
cm
5.50
2.50
4.50
–
Power
%
100
100
100
–
NOTE 1 Only one operating condition per index.
NOTE 2 Data should be entered for “at surface” and “below surface” both in the columns related to TIS or
TIB.
NOTE 3 Information need not be provided regarding TIC for any TRANSDUCER ASSEMBLY not intended
for
transcranial or neonatal cephalic uses.
NOTE 4 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns related
to TIS, TIB or TIC.
NOTE 5 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the column related
to MI.
NOTE 6 The depths z
pii
and z
pii,α
apply to NON-SCANNING MODES, while the depths z
sii
and z
sii,α
apply to
SCANNING MODES.
Содержание ECO 1VET