ECO Series VET Digital Color Doppler Ultrasound System
20
Configuration ECO 1VET
ECO 2VET
ECO 3VET
ECO 3
EXPERT VET
ECO 5VET
Probe
1(standard),
2(option)
1(standard),
2(option)
2(standard)
2(standard)
2(standard)
Maximum
memory
8GB
8GB
8GB
8GB
8GB
HDD
No
Option
Option
Option
Option
Built in battery Option
Option
Standard
Standard
Standard
Compound,
i-Image,
SRA,THI
Standard
Standard
Standard
Standard
Standard
Trapezoidal
No
Standard
Standard
Standard
Standard
PW
Mode
No
Standard
No
Standard
Standard
CFM Mode
No
No
No
No
Standard
Chroma
Standard
Standard
Standard
Standard
Standard
DICOM 3.0
Option
Option
Option
Option
Option
Graphic
Printer
Option
Option
Option
Option
Option
Probe
configuration
Seven:
C3-A,
V6-A, MC3-A,
L7M-A, L7S-A,
MC6-A, R7-A
Seven:
C3-A,
V6-A, MC3-A,
L7M-A, L7S-A,
MC6-A, R7-A
Seven:
C3-A,
V6-A,
MC3-A,
L7M-A, L7S-A,
MC6-A, R7-A
Seven:
C3-A,
V6-A, MC3-A,
L7M-A, L7S-A,
MC6-A, R7-A
Seven:
C3-A,
V6-A, MC3-A,
L7M-A, L7S-A,
MC6-A, R7-A
3.5
Installation Procedures
Note: Please do not turn on the power switch until finishing all the installation and necessary preparation.
3.5.1
Environment Condition
The system should be operated under the following environment.
3.5.1.1
Operation Environment Requirement
Ambient Temperature: 10
℃~
40
℃
Relative Humidity: 30%
~
75%RH
Atmospheric Pressure: 700hPa
~
1060hPa
Strong radiation sources or powerful electromagnetic waves (e.g. electro-magnetic waves from radio
broadcasting) may result in image ghosting or noise. The system should be isolated from such radiation
sources or electromagnetic waves.
To prevent damage to the system, do not use in the following locations:
Exposed to direct sunlight
Subject to sudden changes in temperature
Dusty
Subject to vibration
Near heat generators