![Chison ECO 1VET Скачать руководство пользователя страница 156](http://html.mh-extra.com/html/chison/eco-1vet/eco-1vet_operation-manual_2602722156.webp)
ECO Series VET Digital Color Doppler Ultrasound System
141
Transducer Model: MC5V-A
Operating Mode: PW
Index label
MI
TIS
TIB
TIC
At
surface
Below
surface
At
surface
Below
surface
Maximum index value
0.20
0.80
0.70
–
Index component value
0.74
0.80
0.66
0.70
Acoustic
Parameters
p
r,α
at Z
MI
Mpa
0.49
P
mW
82.00
82.00
–
P
1x1
mW
28.47
28.47
Z
s
cm
2.22
Z
b
cm
2.30
Z
MI
cm
0.01
z
pii,α
cm
0.01
f
awf
MHz
5.32
5.27
5.35
–
Other
Information
prr
Hz
4000.00
srr
Hz
–
n
pps
1
I
pa,α
at z
pii,α
W/cm
2
9.26
I
spta,α
at z
pii,α
or
z
sii,α
mW/cm
2
10.23
I
spta
at z
pii
or z
sii
mW/cm
2
33.21
p
r
at z
pii
Mpa
0.56
Operating
control
conditions
Focus
cm
5.00
5.00
5.00
–
Power
%
100
100
100
–
NOTE 1 Only one operating condition per index.
NOTE 2 Data should be entered for “at surface” and “below surface” both in the columns related to TIS or
TIB.
NOTE 3 Information need not be provided regarding TIC for any TRANSDUCER ASSEMBLY not intended
for
transcranial or neonatal cephalic uses.
NOTE 4 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns related
to TIS, TIB or TIC.
NOTE 5 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the column related
to MI.
NOTE 6 The depths z
pii
and z
pii,α
apply to NON-SCANNING MODES, while the depths z
sii
and z
sii,α
apply to
SCANNING MODES.
Содержание ECO 1VET
Страница 2: ......