![Chison ECO 1VET Скачать руководство пользователя страница 138](http://html.mh-extra.com/html/chison/eco-1vet/eco-1vet_operation-manual_2602722138.webp)
ECO Series VET Digital Color Doppler Ultrasound System
123
Transducer Model: R7-A
Operating Mode: B+C
Index label
MI
TIS
TIB
TIC
At
surface
Below
surface
At
surface
Below
surface
Maximum index value
0.40
0.20
0.20
–
Index component value
0.20
0.20
0.11
0.20
Acoustic
Parameters
p
r,α
at Z
MI
Mpa
1.09
P
mW
16.00
16.00
–
P
1x1
mW
3.92
3.92
Z
s
cm
–
Z
b
cm
–
Z
MI
cm
1.22
z
pii,α
cm
–
f
awf
MHz
7.87
7.87
7.87
–
Other
Information
prr
Hz
–
srr
Hz
47.63
n
pps
1
I
pa,α
at z
pii,α
W/cm
2
–
I
spta,α
at z
pii,α
or
z
sii,α
mW/cm
2
36.65
I
spta
at z
pii
or z
sii
mW/cm
2
55.23
p
r
at z
pii
Mpa
–
Operating
Control
Conditons
Focus
cm
1.00
1.00
1.00
–
Power
%
100
100
100
–
NOTE 1 Only one operating condition per index.
NOTE 2 Data should be entered for “at surface” and “below surface” both in the columns related to TIS or
TIB.
NOTE 3 Information need not be provided regarding TIC for any TRANSDUCER ASSEMBLY not intended
for
transcranial or neonatal cephalic uses.
NOTE 4 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns related
to TIS, TIB or TIC.
NOTE 5 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the column related
to MI.
NOTE 6 The depths z
pii
and z
pii,α
apply to NON-SCANNING MODES, while the depths z
sii
and z
sii,α
apply to
SCANNING MODES.
Содержание ECO 1VET