ECO Series VET Digital Color Doppler Ultrasound System
142
Transducer Model: L7V-A
Operating Mode: B
Index label
MI
TIS
TIB
TIC
At
surface
Below
surface
At
surface
Below
surface
Maximum index value
0.30
0.10
0.10
–
Index component value
0.10
0.10
0.09
0.10
Acoustic
Parameters
p
r,α
at Z
MI
Mpa
0.70
P
mW
16.00
16.00
–
P
1x1
mW
3.21
3.21
Z
s
cm
–
Z
b
cm
–
Z
MI
cm
1.88
z
pii,α
cm
–
f
awf
MHz
7.23
7.21
7.21
–
Other
Information
prr
Hz
–
srr
Hz
30.05
n
pps
1
I
pa,α
at z
pii,α
W/cm
2
–
I
spta,α
at z
pii,α
or
z
sii,α
mW/cm
2
33.23
I
spta
at z
pii
or z
sii
mW/cm
2
55.21
p
r
at z
pii
Mpa
–
Operating
control
conditions
Focus
cm
2.00
6.00
6.00
–
Power
%
100
100
100
–
NOTE 1 Only one operating condition per index.
NOTE 2 Data should be entered for “at surface” and “below surface” both in the columns related to TIS or
TIB.
NOTE 3 Information need not be provided regarding TIC for any TRANSDUCER ASSEMBLY not intended
for
transcranial or neonatal cephalic uses.
NOTE 4 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns related
to TIS, TIB or TIC.
NOTE 5 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the column related
to MI.
NOTE 6 The depths z
pii
and z
pii,α
apply to NON-SCANNING MODES, while the depths z
sii
and z
sii,α
apply to
SCANNING MODES.