MAINTENANCE
Only qualified personnel may carry out maintenance on the equipment. Technicians must be familiar
with the model, use and maintenance of the equipment.
Check the equipment whenever it has undergone a "rough treatment" (e.g.: fall, liquid spill, etc.) and
whenever any doubts arise concerning its safety.
5.1
Periodic maintenance
Periodic maintenance must be carried out at least once every 12 months, following the table below:
WARNING
when safety is at risk, immediately turn off the equipment and do not use it until the danger has
been eliminated.
any significant work carried out to ensure the couch is safe for use must be recorded in the
relevant form (attached).
FREQUENCY
OPERATION
Yearly
General clean of the equipment
Visual and functional inspection of the equipment
Frame and mechanical parts: check screws and bolts to ensure they are
tight; pin and bushing clearance; check for any damage or deformation
Check that the adjustable sections are stable and do not collapse when the
control lever is released
Check spring control cables and respective sheaths
Check the wheels are firmly secured and free to rotate; clean the tread
Check control box, actuators and controls to see if connectors are fully
plugged in (electric versions)
Check insulation on electrical cables for damage (electric versions)
Make sure that during electric operation no unusual vibrations or sounds are
produced that may signal defects in the actuators and/ or mechanisms.
After the above operations, complete inspection by carrying out a full lifting cycle using the maximum
safety working load.
5.2
Extraordinary maintenance
The extraordinary maintenance must be performed by skilled technicians with knowledge of the
contents of this instruction booklet.