Instruction Manual
SMART-LIFT Electric Ceiling Lifts
16
Extend/Retract for Momentary or Latching Contacts
These terminals can be used with any latching contacts, momentary
contacts, or a wall switch
NOTE: The connection between the ground (#9) and any other ter-
minal connection must be broken (open) before completing the
next circuit.
Wire for contact operation (Figure 18) as follows:
1.
Install a wire between terminal #9 (shown) and contacts or
switch (see Figure 19).
2.
Connect power source
3.
For extend, complete the circuit to terminal #4 (see Figure 20).
4.
Make sure circuit to terminal #4 is open and, for service
extend, complete the circuit to terminal #11.
5.
Make sure circuit to terminal #11 is open and, for retract, com-
plete the circuit to terminal #3.
Figure 20. Contacts Connection
Figure 19. Contacts Connection
Use Category II Wiring
Figure 18. Contacts Wiring