3.2.4 Single Pulse mode
3.2.5 Burst mode
This mode lets you set the duration of a single pulse. The Single Pulse mode
delivers a pulse, the duration of which is defined by the operator each time
the footswitch is pressed. It is possible to select a duration of 20 ms to a
maximum of 1000 ms by pressing the square at the center of the screen.
Pressing “Start” and applying pressure to the footswitch will activate a single
pulse, which will be delivered for the pre-specified time frame. Continuing
to press the footswitch will deliver an identical pulse. The lower part of the
screen shows the actual power delivered, the time, the amount of energy
transferred and the temperature of the laser head.
This mode lets you set the pulse frequency and the duration of each pulse.
Selecting the pulse duration <500ms allows you to change the frequency
(10Hz max) by making the appropriate selections. After setting the
parameters of frequency and pulse duration by clicking on “laser power” you
can change the working power. By removing the handpiece from its holder,
putting on the applicator, pressing “Start” and acting on the pneumatic
footswitch, you can begin the therapy session. The lower part of the
screen shows the actual power delivered, the time, the amount of energy
transferred and the temperature of the laser head.