17
EN
3.2.2 Interrupted mode
3.2.3 Pulsed mode
To start the treatment choose Interrupted emission mode, set power
parameter by pressing “laser power” and energy quantity by pressing “Joule”.
The duty cycle can be modified. Take out the handpiece from its holder, put
on the handpiece spacer and press the footswitch to activate laser emission.
The lower part of the screen shows the actual power delivered, the time, the
amount of energy transferred and the temperature of the laser head.
The pulsed mode is the most commonly indicated for the acute phase
treatment, optimizing the anti-inflammatory action and increasing the bio-
stimulation while minimising thermal impact. This mode enables the laser
emission optimization depending on the skin type.
By choosing the duration of the impuls < 500ms it is possible to change the
frequency by selecting the appropriate option on the screen.
After setting the parameters of frequency and pulse duration by clicking
on “laser power” you can change the working power, while clicking “joules”
in the upper right of the screen allows you to set the amount of energy to
be transferred. By removing the handpiece from its holder, putting on the
applicator, pressing “Start” and acting on the pneumatic footswitch you can
begin the therapy session.
The lower part of the screen shows the actual power delivered, the time,
the amount of energy transferred and the temperature of the laser head. To
change the parameters of time and energy you need to press the square at
the top of the screen.