Refer to the table below for troubleshooting. For problems of any other type
please contact the DJO Assistance Centre.
6.2 Troubleshooting
and periodical
maintenance
Problem
Probable Cause
Solution
Intelect HPL 7 / HPL15 does
not switch on.
A) No power supply
B) Blown fuse
C) Power cable damaged or plug
not fully pushed in
A) Check power supply
B) Change the fuse
C) Change the damaged cable or
connect correctly
The machine does not
emit energy
A) Exit circuit interrupted
B) Emergency stop button pressed
C) Safety interlock open
A) Contact the DJO Assistance Centre
B) Press and turn clockwise
C) Check the interlock
The laser turns on but the
display shows nothing
A) Probable display fault
A) Contact the DJO
Assistance Centre
Pressing a button does not have
the desired effect
A) Probable button fault
B) Probable control circuit fault
A) Contact the DJO
Assistance Centre
The calibration check
shows ‘FAULT’
A) Probable optical fibre breakage
B) Degraded laser emission
C) Calibration check system
deteriorated or dirty
A) Clean the laser protection window
B) Clean the calibration unit
C) Contact the DJO
Assistance Centre
Checking of the electronic circuits must be carried out by personnel who
are highly qualified and authorized by a document from the manufacturer.
Therefore the manufacturer must be contacted to maintain the conditions
of the guarantee.
We would be happy to answer any questions you may have about our
products and services.
Please contact your local dealer, or your corresponding DJO Global site.
DJO Global sites are listed on back page.