31
EN
The circuitry is generally oversized and well ventilated, which avoids
overheating and associated problems.
The recommended operating temperature is between 10 and 30 °C, with
humidity between 30 and 75% and pressure between 700 and 1060 hPa.
When used in a clinic or hospital, in a normal non-sterile environment,
Intelect HPL 7 and HPL15 has no particular protection against accidental
spillage of liquids; occasional splashes or sprays of water or gel do not create
problems due to the distance between circuit boards and ventilation grills.
The recommended environments for use are: hospital, clinic and
private center.
The recommended environments for use have to have the following
environmental and electrical requirements:
-) electrical system that respects the standards of the installation country
-) no reflective surface like mirror
-) laser attention symbol
-) red light outside the laser therapy room
-) labels that underline that it is necessary to use laser protection glasses
during the therapy
-) interlock system connected to the door of the laser therapy room
For the disposal of the machine at the end of its life, follow WEEE
requirements on page 4.
Содержание Intelect HPL15
Страница 1: ...1 EN User Manual Intelect HPL7 and HPL15 ...
Страница 38: ...38 EN 10 EMC tables ...
Страница 39: ...39 EN ...
Страница 40: ...40 EN ...
Страница 41: ...41 EN ...