Technical Specifications
Listed here are the specifications for the Intelect
®
Legend Ultrasound.
Dimensions: 8.25” x 11” x 2.5”
Weight: 2
lb.
Applicators: 2, 5, & 10 cm²
Power:
90V-264V, 47-63Hz
2 cm² Applicator
Frequency (Mhz):
1.0 MHz, 3.3 MHz (all +/- 5%)
Power (watts):
0.2 watt to 4 watts
ERA
(cm²):
1.8
cm²
Effective Radiating Area: 1.8 cm² +/-0.4 cm²
Maximum beam
non-uniformity ratio:
6.0:1
Beam
type:
Collimating
5 cm² Applicator
Frequency (Mhz):
1.0 MHz, 3.3 MHz (all +/- 5%)
Power (watts):
0.4 watt to 10 watts
ERA
(cm²):
4.0
cm²
Effective Radiating Area: 4.0 cm² +/-1.0 cm²
Maximum beam
non-uniformity ratio:
6.0:1
Beam
type:
Collimating
10 cm² Applicator
Frequency (Mhz):
1.0 MHz, 3.3 MHz (all +/- 5%)
Power (watts):
1 MHz: .1 watt to 20 watts, 3.3 MHz: 1-10 watts
ERA
(cm²):
8.5
cm²
Effective Radiating Area: 8.5 cm² +/-1.5 cm²
Maximum beam
non-uniformity ratio:
6.0:1
Beam
type:
Collimating
Duty Cycle
Pulsed:
10%, 20%, 50%
Continuous: 100%
Pulse Duration:
5 msec +20% (50% duty cycle, pulsed mode)
2 msec +20% (20% duty cycle, pulsed mode)