10. Push the START button. The unit will hold the settings in memory. The display will
change to CAL2.
11. If you are satisfied with the calibration, push START.
12. The unit will beep to indicate that the new calibration values have been stored in the
applicator.
13. The display will change to CAL0 and is now ready for the next calibration.
14. After both frequencies have been calibrated and before removing the applicator
head from the power meter, check the accuracy of the power calibration. Cycle the
power to the unit off and then back on. Select continuous duty cycle and increase the
output to the settings listed. Verify that the power indicated on the UPM-DT-10 power
meter is within the indicated range. Check power at both 1 MHz and 3.3 Mhz.
Head size
Power level
Output range
2 cm
2
2 watts
1.72-2.28
4
watts 3.44-4.56
5 cm
2
5 watts
4.3-5.7
10
watts
8.6-11.4
10 cm
2
5 watts
4.3-5.7
10
watts
8.6-11.4
15
watts
12.9-17.1
20
watts
17.2-22.8
15. Calibration complete. If you are not satisfied with the calibration and wish to start
over, turn the machine OFF and begin at step one listed above.
.
Innovation starts here
TM
38
Intelect
®
Legend Series Service Manual