5. Press DUTY CYCLE until Continuous is illuminated, noting that each LED is
functioning. Use the INTENSITY up button to set the output to 2.0W/cm
2
.
The fan should begin running. Press START. By applying a small amount of
water to the transducer, verify that there is ultrasound output.
6. Observe that the treatment time counts down on the LEDs. Press PAUSE.
The output should stop, the fan should stop and the treatment timer should
stop counting down. Press START and the output should resume. Press
STOP and verify that the output stops, the treatment timer is reset back to the
default, the output display is reset back to 0.0 and the fan stops.
7. Use the TIME up button to verify that the treatment time will not go above 15
minutes. Use the DOWN button to set the time back to 1 minute. Select 10%
Duty Cycle using the DUTY CYCLE button, increase the intensity to maximum
and press START. While the unit is running, press the DISPLAY button and
verify that the displayed power changes from 2.0 W/cm
2
to 10.0 Watts
(assuming a 5cm applicator is attached). Allow the unit to run until treatment
time runs out. Two beeps should be heard followed by another two beeps
and then the display will clear back to the default.
8. Testing complete.
Intelect
®
Legend Applicator
Calibration Procedure
1. Put unit into calibration mode. Turn the unit ON while holding down the Intensity UP
button. It is not necessary to hold the button down after the initial beep sound.
2. The Min/Sec display should read CAL0 indicating that the machine is in calibration
mode and at the first calibration step.
3. Place the applicator into the Ultrasound power meter.
4. Select the calibration frequency using the FREQUENCY button (1MHz or 3MHz).
5. Set the Head Size using the DUTY CYCLE button.
10% = 1cm applicator
20% = 2cm applicator
50% = 5cm applicator
Continuous = 10cm applicator
6. Once Applicator size and Frequency are chosen, push the START button. This will
begin the scan and calibration. The zero in CAL0 will flash while the scan is in progress.
Do not disturb the machine while it is scanning.
7. When the scan is complete the display will change to CAL1.
8. Re-zero the Ultrasound power meter.
9. Using the Intensity UP button, increase the output until the power meter reads 4
(four) watts.
Innovation starts here
TM
37
Intelect
®
Legend Series Service Manual