2.
Premod
: With the oscilloscope probe on channel 2, press Channel 2
button. Press Premod and observe: Channel 2 Premod. Adjust output
with Intensity arrow to 50ma and press Start. Main screen should display
Channel 2 Premod with timer counting down from 15:00 and 50ma as
output with Ch1 - Avail below. Verify waveform is similar to that below
using the oscilloscope. Press Stop to reset.
Premod Waveform; ‘scope set to 10V/div, 1mSec/div
3.
HiVolt
: Press HiVolt and Main Menu should display at top: Ch2: High
Volt. Time should be set to 20:00 and Int should now read 0v. Press Enter
to select Probe. Output must display 0v. Press Enter again to select Pads.
Increase Output with the Intensity up button to 500. Press Start and verify
that the output waveform is similar to that shown below. Press Enter and
use Main Menu down button to move to Polarity. Toggle polarity from
positive to negative and observe that the waveform changes as shown
below. Press Stop upon successful completion.
Positive
Negative
Hi-Volt Waveforms; ‘scope set to 50V/div, 20uSec/div
4.
Russian
: Press Channel 1 button. Press Russian & observe: Channel 1
Russian . At Mode, press Enter and select CoContract and press Enter
again. Main Menu heading must say: Ch1-2: Russian. Move down to
Cycle Time with the arrow keys and press Enter. Select Continuous and
press Enter again. Increase Intensity with the arrow key to maximum
100ma and press Start. Verify on Main screen that both Channels reach
100 by pressing the Channel 1 and 2 button. Verify that the output