waveform on both channels is similar to that shown below using the
oscilloscope.
Russian Waveform; ‘scope set to 50V/div, 5mSec/div
Press pause to pause the operation and then press the Up and Down Time
buttons to confirm that the time changes. Set the treatment time to 1 minute
and push Start. Upon completion, you should hear 2 beeps, a pause, and
then 2 more beeps to indicate end of treatment. The output must be at zero
at this time.
C. Screen Intensity Adjustment
1. Hold Channel 1 and Intensity up or down button to adjust the screen for
proper display.
Remove power; Test is complete!
Intelect
®
Legend US
Calibration Procedure
Note: All caps indicate a front panel button.
1. Plug the applicator and appropriate power supply into the unit. Plug the
power supply into an appropriate outlet and turn the unit on. It should beep
and the LEDs should illuminate, then fill with default time, duty cycle, and
frequency.
2. Unplug the applicator. The message ‘noHd’ should appear in the treatment
time display. Plug the applicator back in, press a key and observe that the
message disappears.
3. Press HEAD WARM. The fan should start and output will be displayed on the
LEDs. Pressing it a second time will stop the fan and the output will return to
0.0.
4. Press the FREQUENCY button and verify that 1 MHz and 3 MHz LEDs are
illuminated. Leave 1 MHz selected.
Innovation starts here
TM
36
Intelect
®
Legend Series Service Manual