Intelect Legend Stim Error Messages Explained
Below is a list of the messages displayed on the LCD screen when an
error is detected, as well as a brief explanation of the source of the message.
“Channels Needed for This Mode are Busy”
Either no channels are open to run a new signal, or the dedicated
channel for HiVolt is already being used.
“Over Current Fault on Channels X-X”
Where X is a channel number. Indicates that a severe overcurrent
was detected on the specified channel(s).
Intelect Legend US Error Messages Explained
Below is a list of the codes displayed on the LEDs when an error is
detected.
HHH
The applicator has reached a temperature that is too high, and the
US output has been stopped.
noHd
No ultrasound applicator detected.
E03
An attempt to set up an address on the onewire device failed.
E04
An attempt to read data from the ultrasound applicator failed.
E05
An attempt to program the ultrasound applicator failed. This only
occurs during calibration.
E06
The ultrasound applicator memory is full. This only occurs during
calibration.
E07
The memory device in the ultrasound applicator is not the proper
type.
E08
The ultrasound applicator is uncalibrated.