1. First adjust an angle of 90° for
external rotation and deactivate
motor B as follows:
a. Press the MENU key on the
programming unit to access level 1
(M1).
b. Press the
parameter key
for
external rotation
and move
the carriage to the rotation position
of 90°, using the
+ / - keys
.
c. Press the
MENU key
on the
programming unit to access level
3 (M3).
d. Press the
Motor B ON/OFF
parameter key to activate
the parameter.
e. Deactivate motor B by pressing
the parameter key Motor B
ON/OFF again or the "–" (minus)
key. The check mark in the circle
next to the symbol must have
disappeared.
f. Save the entered values by press
the
STOP key
.
2. Complete the mechanical settings
and adjust the carriage to the patient
as follows:
a. Adjust an angle of 90° for horizon-
tal adduction/abduction
†
, using
indexing knob (12).
b. Secure and hold the motion
element at the holding tube for
length adjustment and open the
locking screw (14) for height
adjustment
Œ
. Select a height at
which the pivot points of motor A
and the shoulder are at the same
level.
c. Now open clamping lever (18) for
adjustment of the upper arm
length
Å
, locking screw (20) for
adjustment of the elbow angle
è
and locking screw (28) for adjust-
ment of the forearm rest inclina-
tion and set the motion element to
a position that is comfortable to
the patient and suitable for the
therapy session . When done,
tighten all screws and levers.
d. Open clamping lever (26), adjust
the forearm length
ì
to the
patient and close the clamping
lever.
3. In the next step, program the eleva-
tion range of motion as follows:
a. Press the
MENU key
on the
programming unit to access level
1 (M1).
b. Next set the desired range of
motion for adduction and abduc-
tion by first pressing the
parame-
ter key
for
adduction
and
set the values with the
+/-
keys.
c. Set the
abduction
value in
the same way.
4. Save the settings with the
STOP key
,
set the carriage to the starting
position with the
START key
and
initiate therapy by pressing the
START key
again.
Note!
Having programmed the ranges of
motion, you can program additional
options, such as pause or speed.
Note!
- For an exclusive elevation protocol,
motor B for the rotation movement
must be turned off as described in
steps c to e below.
- Having programmed the ranges of
motion, you can program additional
options, such as pause or speed.
5.4.3 Isolated elevation
(flexion)
26
OPTIFLEX S Gebr DV2 e 709 25.09.2007 11:16 Uhr Seite 26
Содержание OptiFlex SHOULDER CPM
Страница 3: ...3 D A E B F C Illustraciones Figures OPTIFLEX S Gebr DV2 e 709 25 09 2007 11 16 Uhr Seite 3...
Страница 37: ...37 English OPTIFLEX S Gebr DV2 e 709 25 09 2007 11 17 Uhr Seite 37...
Страница 73: ...73 Espa ol OPTIFLEX S Gebr DV2 sp 708 24 09 2007 18 57 Uhr Seite 73...
Страница 74: ...Notas 74 OPTIFLEX S Gebr DV2 sp 708 24 09 2007 18 57 Uhr Seite 74...
Страница 75: ...Notas 75 Espa ol OPTIFLEX S Gebr DV2 sp 708 24 09 2007 18 57 Uhr Seite 75...