Follow these steps to program pauses:
• Access menu level 2, then press the
parameter key to select the special
function .
• The display indicates the symbol for
the special function and the selected
pause durations.
top: abduction/external rotation
pause
bottom: adduction/internal rotation
pause (see display image)
The pause duration for
abduction/external rotation appears
in a dotted frame.
• To change the abduction/external
rotation pause, press the "+" or "
-
"
key. Select a duration of 10 seconds,
for instance.
• Press the
parameter key
again
when done. The dotted frame
automatically moves to the line
below, indicating that the adduc-
tion/internal rotation pause can be
adjusted.
The
parameter key
is only used to
toggle between these two functions.
• To modify the value, press the "+"
(plus) or "-" (minus) key. Select a
duration of 5 seconds, for instance.
• Save the programmed values by
pressing the
STOP
key, then press
the
START
key to begin the treat-
ment.
■
Therapy timer
Default setting
is
continuous
operation
.
A clock symbol in the upper right-
hand corner of the display identifies
the continuous mode of
operation.The clock indicates the
elapsed therapy time.
In the continuous mode, the device
must be stopped with the
STOP
key.
However, the therapy timer can be
set in
steps of 1 minute to any
value between 1 and 300 minutes
.
At the end of the therapy time, the
device automatically stops in the
starting position (maximum adduction
position / halfway between internal
and external rotation).
In this case, a circle replaces the
clock symbol. The circle fills as the
therapy time progresses.
■
Speed
The speed can be adjusted between
1 % and 100 % in steps of 1 %.
100 % is equivalent to 230°/minute
Default: 100 %
■
New patient
With this function, the CPM device
will move to the home position.
• Select the function and press the
START
key:
– The carriage moves to the home
position.
– The programmed treatment
parameters are deleted.
– All values stored on the chip card
are deleted.
– The device stops halfway between
the abduction/adduction and
internal/external rotation position.
The "New patient" function (home
position) selects the following
settings:
22
OPTIFLEX S Gebr DV2 e 709 25.09.2007 11:16 Uhr Seite 22
Содержание OptiFlex SHOULDER CPM
Страница 3: ...3 D A E B F C Illustraciones Figures OPTIFLEX S Gebr DV2 e 709 25 09 2007 11 16 Uhr Seite 3...
Страница 37: ...37 English OPTIFLEX S Gebr DV2 e 709 25 09 2007 11 17 Uhr Seite 37...
Страница 73: ...73 Espa ol OPTIFLEX S Gebr DV2 sp 708 24 09 2007 18 57 Uhr Seite 73...
Страница 74: ...Notas 74 OPTIFLEX S Gebr DV2 sp 708 24 09 2007 18 57 Uhr Seite 74...
Страница 75: ...Notas 75 Espa ol OPTIFLEX S Gebr DV2 sp 708 24 09 2007 18 57 Uhr Seite 75...