21
English
■
Internal rotation
Maximum value:
- 90 degrees
■
External rotation
Maximum value:
90 degrees
■
Pauses
Pauses occur at each programmed
maximum value.
These are the two pause points:
- Maximum adduction value simulta-
neous with maximum internal
rotation value
- Maximum abduction value simulta-
neous with maximum external
rotation value
Pauses are adjustable in steps of 1
second between 0 and 30 seconds.
Default: no pause
LEVEL 2:
the carriage will first move to the
maximum adduction position and to
the position halfway between
internal and external rotation. Then
the carriage will stop, allowing the
patient to be positioned. When the
START key
is pressed again (start
of therapy), the carriage will first
move to the maximum internal
rotation position. Motor A (abduc-
tion/adduction) stops during this
procedure. Once the maximum
internal rotation position has been
reached, both motors (A and B) will
simultaneously move to the maxi-
mum abduction and external
rotation position. After reaching this
position, the cycle restarts:
maximum adduction and maximum
internal rotation.
• The elevation movement (flexion) is
programmed by means of the
abduction/adduction values (for
mechanical adjustment instructions,
refer to section 5.6 "Examples of
use").
• Horizontal adduction/abduction is
adjusted manually. These are
possible settings:
Horizontal adduction:
120°
Horizontal abduction:
0°
Note!
• The programmed values and the
values (degrees) measured on the
patient may deviate slightly.
• To ensure a physiological motion,
the carriage simultaneously moves
to the following positions in the
synchronized mode:
- Maximum abduction value simulta-
neous with maximum external
rotation value
- Maximum adduction value simul-
taneous with maximum internal
rotation value
• To ensure that the movements you
program are physiological in nature,
we recommend to program the
carriage in the following order:
1. Maximum adduction value
2. Maximum internal rotation value
3. External rotation
(to relieve the shoulder)
= 0°
5. Maximum abduction value
6. Maximum external rotation value
• In the synchronized mode, after
activation of
START
OPTIFLEX S Gebr DV2 e 709 25.09.2007 11:16 Uhr Seite 21
Содержание OptiFlex SHOULDER CPM
Страница 3: ...3 D A E B F C Illustraciones Figures OPTIFLEX S Gebr DV2 e 709 25 09 2007 11 16 Uhr Seite 3...
Страница 37: ...37 English OPTIFLEX S Gebr DV2 e 709 25 09 2007 11 17 Uhr Seite 37...
Страница 73: ...73 Espa ol OPTIFLEX S Gebr DV2 sp 708 24 09 2007 18 57 Uhr Seite 73...
Страница 74: ...Notas 74 OPTIFLEX S Gebr DV2 sp 708 24 09 2007 18 57 Uhr Seite 74...
Страница 75: ...Notas 75 Espa ol OPTIFLEX S Gebr DV2 sp 708 24 09 2007 18 57 Uhr Seite 75...