20994124 REV A
43
constantly monitor, charge, and sense the current in its
battery circuit. With its main power connected, the GUI
will constantly produce a 27 VDC trickle charge to keep
the batteries fully charged.
INTERNAL BATTERY REPLACEMENT PROCEDURE
1. Make sure the power supply and the battery backup are
disconnected.
2. Remove Cowling (see
Figure
1) from the Liquefier, by
removing the four screws that fasten the shroud to the
top of the freezer.
3. Remove top cover off the Liquefier, unscrewing the
screws that hold the top cover of the liquefier in place
(see Figure 26).
4. Remove the foam insert that sits over the fan and battery
backup.
5. Remove the four Phillips screws holding the backing
plate (Figure 27).
6. Replace the new Battery Backup (PN
21183303).