20994124 REV A
39
Table 4 : Preventative Maintenance Procedures
Item
No
Action
Procedure
1
Level
Verification
1. Liquid Level can only be verified via the GUI.
2
Lid Inspection
1. Remove the lid slowly from the MVE Fusion.
2. Inspect lid for any damage and replace parts if necessary
3
Lid
Thaw
Procedure
1. Remove the lid slowly from freezer.
2. It is recommended that the freezer opening be covered
with a spare lid or in another non-airtight manner to
prevent moisture from entering the storage space and to
minimize the top box temperature change while the lid is
open.
3. Allow lid to sit at room temperature for approximately 30
minutes. (Caution-the lid open alarm will sound, press
alarm mute to temporarily silence the alarm)
4. Once thawed, thoroughly dry lid, cork, and liner.
5. Inspect lid for damage and replace parts if necessary.
4
Folding
Step
Inspection
1. Verify that hinges and free of cracks and all connections
are secure.
2. Check that the anti-slip strips on the steps are in good
condition and replace if necessary (PN 4810179).
3. Ensure the step locking strap is able to securely hold the
steps in their folded position. If the pivot bolts
continuously loosen, apply thread locker (PN 11087674)
and retighten.
Содержание MVE Fusion
Страница 1: ...20994124 REV A 1 MVE FUSION Technical Manual ...
Страница 8: ...20994124 REV A 8 Figure 2 Cross Section of the MVE Fusion Freezer ...
Страница 16: ...20994124 REV A 16 Figure 9 Left to Right Vent Overfill Valve Relief Valve and Fill Valve ...
Страница 35: ...20994124 REV A 35 Figure 24 a Format of the Data logs screen b Format of the Data file ...
Страница 50: ...20994124 REV A 50 NOTES ...