Operation
5.1 Basic Operation
The MVE Fusion is a “self-sustaining” cryogenic storage freezer which normally requires
limited user intervention once it has been installed at the customers site and Chart (or an
authorized distributor) performs the first fill. However, there are circumstances (power
outages, for instance) when the system will require attention. There are alarms and other
information provided by the GUI (graphical user interface) to convey important messages
and warnings to the user. Understanding and familiarizing yourself with the Fusion freezer
and the its GUI is recommended for the user for the smooth functioning of the system.
The MVE Fusion uses an Android tablet with a custom-built software application to
monitor the freezers operation.
CAUTION: The user should never operate the Fusion freezer without the
Android running the custom built software application and should never
exit the application during operation.
5.1.1 Starting the Fusion Freezer Safely
This section of the manual will detail the functions and features of the Android application
and demonstrates how to access and monitor the freezer operation.
CAUTION: The Fusion must have LN2 inside the pressure vessel before
electrical power is supplied to it (see Section 4.1.1 for how to fill the
Fusion).
1. Connect the included electric power cord to the back of the unit.
2. Switch the System Power and the Battery Backup power switches (see Figure 19)
to the “On” position (I).
Содержание MVE Fusion
Страница 1: ...20994124 REV A 1 MVE FUSION Technical Manual ...
Страница 8: ...20994124 REV A 8 Figure 2 Cross Section of the MVE Fusion Freezer ...
Страница 16: ...20994124 REV A 16 Figure 9 Left to Right Vent Overfill Valve Relief Valve and Fill Valve ...
Страница 35: ...20994124 REV A 35 Figure 24 a Format of the Data logs screen b Format of the Data file ...
Страница 50: ...20994124 REV A 50 NOTES ...