20994124 REV A
41
Figure 26: After removing the Screws on the liquefier
4. Take a cloth and wipe away any dust or debris off the
two fans.
8
Relief
Valve
Replacement
CAUTION
: Ensure that the LN2 supply valve is
closed and the plumbing assembly is vented before
removing the relief valve.
1. If equipped with a relief valve deflector, loosen the
deflector clamp and slide off the deflector.
2. Loosen the relief valve and remove it from the plumbing
assembly. Be sure to support the attachment tube with
wrench to prevent damage from twisting.
3. Install new relief valve (PN 1810032) applying new
Teflon tape if needed. Ensure relief valve is rated to 50
PSI (3.4 bar).
CAUTION
: Installing a relief valve with a different
pressure rating could prevent proper operation and