Part Number: 7200-1042
Page 1 of 1
TITLE: Procedure, CHA-Control Software Installation and/or Re-Installation
Instrument: Model 7200 CHA (Cement Hydration Analyzer)
Revision Date Revised
By Description
A 10/16/2007
DAH/WJW ECN
T1161
This procedure is for installation of the CHA-Control software onto a new computer, for re-installation of the
software on the same computer, or for upgrading the software with a new version. These instructions assume that
Microsoft Windows XP is installed and that you are logged in with “Administrator” privileges.
Steps:
(1) If you are upgrading the software, or re-installing it on the same computer, and have calibration data for
the CHA instrument that you wish to re-use, then first copy the “globals” file from the folder
C:\Program
Files\Chandler Engineering\CHA
to another folder. This file contains the calibration constants and
settings, and other information used by the software.
(2) Install the CHA-Control software by running
setup.exe
from the root folder of the software distribution CD.
Follow the on-screen instructions; restart the computer if instructed to do so.
(3) Install the support files by running
setup.exe
from the
9000Utils
folder on the software distribution CD.
Follow the on-screen instructions; restart the computer if instructed to do so.
(4) If you previously copied the “globals” file to another folder in Step 1 above, now copy this file back to the
folder
C:\Program Files\Chandler Engineering\CHA
.
(5) Set up the network interface (Ethernet, LAN) card of the computer to communicate with the Wago I/O
Ethernet controller (part of the bank of data-acquisition modules inside the CHA instrument cabinet):
a. First, navigate to the “Network Connections” window in Windows XP. This may be done in several
ways:
i.
“Network Connections” may appear on the Start… Settings menu.
ii. Click “My Network Places” from the Start menu, and select “View network connections” from the
“Network Tasks” pane on the left.
iii. Right-click on the network connection icon located in the System Tray of the Taskbar, and select
“Open Network Connections”.
b. Right-click on the appropriate Network Connection (“Local Area Connection”) and select “Properties”.
The Properties window will appear.
c. Under the General tab, select the “Internet Protocol (TCP/IP) icon in the “This connection uses the
following items:” list, and click on the “Properties” button. The “Internet Protocol (TCP/IP) Properties”
window will appear.
d. Select “Use the following IP address:”.
e. Enter “10.1.1.1” in the “IP address” field.
f.
Enter “255.0.0.0” in the “Subnet mask” field.
g. Click the “OK” button to close the “Internet Protocol (TCP/IP) Properties” window.
h. Click the “OK” button to close the “Connection Properties” window.
i.
Close the “Network Connections” window.
(6) The CHA-Control software should now be installed.
(7) Click on the “CHA-Control” icon located on the desktop to run the software. You can also go to
Programs… CHA… CHA-Control.
Содержание AMETEK 7200
Страница 10: ......
Страница 16: ...2 2 SECTION 2 OPERATING INSTRUCTIONS Figure 1 Front Panel...
Страница 38: ......
Страница 48: ...5 2 SECTION 5 TROUBLESHOOTING GUIDE This page is intentionally left blank...
Страница 50: ...6 2 SECTION 6 REPLACEMENT PARTS This page is intentionally left blank...
Страница 52: ...6 2 SECTION 6 REPLACEMENT PARTS This page is intentionally left blank...
Страница 65: ...CHANDLER ENGINEERING...
Страница 66: ......
Страница 76: ...cut out postcards on dotted lines...