ASX-1400 Autosampler Operator’s Manual
Using the Autosampler
5–8
rinse agents prevents degradation and failure of the flow path
components.
To flush the rinse station and flow path, complete the following steps:
1)
Insert the rinse uptake tubing into a deionized water source.
Note:
If you are flushing the rinse system during initial startup, first use a 2%
nitric acid solution as the rinse agent, followed by deionized water.
2)
Run the rinse solution through the rinse station and flow
path for 5 to 10 minutes.
Once you flush the rinse system, you can proceed with the sampling
sequence or drain the rinse system as part of the shutdown
procedure. For information about running the sampling sequence,
see “Starting the Autosampler” earlier in this chapter. For more
information about draining the rinse system, see the previous
section, “Shutting Down the Autosampler.”
Содержание ASX-1400
Страница 1: ...ASX 1400 Autosampler Operator s Manual ...
Страница 12: ...Contents ...
Страница 17: ...Preface ...
Страница 23: ...1 Introduction ...
Страница 31: ...2 Preparing for Installation ...
Страница 35: ...3 Installing the Autosampler ...
Страница 39: ...ASX 1400 Autosampler Operator s Manual Installing the Autosampler 3 5 Figure 3 2 Mounting of Z Drive Assembly ...
Страница 43: ...ASX 1400 Autosampler Operator s Manual Installing the Autosampler 3 9 Figure 3 2 Pumped Drain Arrangement ...
Страница 52: ...4 Verifying Installation ...
Страница 60: ...5 Using the Autosampler ...
Страница 68: ...6 1 6 Maintaining the Autosampler ...
Страница 76: ...7 Troubleshooting the Autosampler ...
Страница 81: ...8 Appendix A ...