ASX-1400 Autosampler Operator’s Manual
Verifying Installation
4–7
12)
Click OK on the COM1_Test Properties box.
13)
Type HOME at the cursor in the upper left of the main
Hyperterminal
®
screen and press Enter.
The autosampler resets, with the sample probe moving out and back
into the home position. If the autosampler does not reset, see
Chapter 7, “Troubleshooting the Autosampler.”
Checking the Autosampler Components
The following autosampler components may be damaged during
shipping or installation: the sample probe, the peristaltic pump tubing,
and the rinse station and tubing. It is important that you check these
components for damage before you operate the autosampler. To do so,
complete the following steps:
1)
Shut down and unplug the autosampler.
2)
Visually inspect the sample probe, peristaltic pump tubing,
and rinse station and tubing for leaks or signs of damage.
If you detect a leak or other damage to an autosampler component,
you must replace it. For more information, see the appropriate
section in Chapter 6, “Maintaining the Autosampler.”
Testing the Sample Probe
The sample probe must descend into the center of each sample vial to
ensure satisfactory sample uptake. Shipping or rough handling can
disturb the autosampler’s cabinet-to-base alignment. If it is incorrectly
aligned, the sample probe will not function properly. It is therefore
important to test the sample probe before you actually run samples with
the autosampler.
Содержание ASX-1400
Страница 1: ...ASX 1400 Autosampler Operator s Manual ...
Страница 12: ...Contents ...
Страница 17: ...Preface ...
Страница 23: ...1 Introduction ...
Страница 31: ...2 Preparing for Installation ...
Страница 35: ...3 Installing the Autosampler ...
Страница 39: ...ASX 1400 Autosampler Operator s Manual Installing the Autosampler 3 5 Figure 3 2 Mounting of Z Drive Assembly ...
Страница 43: ...ASX 1400 Autosampler Operator s Manual Installing the Autosampler 3 9 Figure 3 2 Pumped Drain Arrangement ...
Страница 52: ...4 Verifying Installation ...
Страница 60: ...5 Using the Autosampler ...
Страница 68: ...6 1 6 Maintaining the Autosampler ...
Страница 76: ...7 Troubleshooting the Autosampler ...
Страница 81: ...8 Appendix A ...